36
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The ACRL-9100 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
RF emission
CISPR 11
Class B
Class A
Complies
Guidance and manufacture’s declaration – electromagnetic immunity
The ACRL-9100 is intended for use in the electromagnetic environment specified below.
The user of ACRL-9100 should assure that it is used in such an environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floor are covered
with synthetic material, the
relative humidity should be
at least 30%.
Immunity test
IEC 60601 test level
Compliance level