50
From measurement mode:
1.
Press the MODE key
to select the
mg/l (ppm) measurement mode.
2.
Press the SETUP key
to enter SETUP
mode.
3.
Press the
▲
and
▼
keys
to scroll
through subgroups until you view
parameter P3.0 in the lower display.
See figure
4.
Press the ENTER key.
The upper
display shows the probe slope.
See figure
6.
Press the ENTER key.
The upper
display shows the 100 % saturation mV
value.
See figure
7.
Press the ENTER key.
The upper display
shows the 0 % saturation mV value.
See figure
8.
Press the ENTER key
to move back to
subgroup P3.0. If you want to return to
measurement mode, press CAL/MEAS.
mg/l (ppm) mode—
P3.0: Electrode properties
These “view only” parameters show the electrode properties for diagnostic purposes:
probe slope:
shows the calibration slope of the probe. This mode displays slope
from 0.5 to 1.999 % of slope (1.0 = 100%).
100% saturation mV value:
shows the sensor’s mV output corresponding to
100% saturation.
0% saturation mV value:
shows the sensor’s millivolt output corresponding to
0% saturation.