42
From measurement mode:
1.
Press the MODE key
to select the
% saturation measurement mode.
2.
Press the SETUP key
to enter SETUP
mode.
3.
Press the
▲
and
▼
keys
to scroll
through subgroups until you view
parameter P4.0 in the lower display.
See figure
4.
Press the ENTER key.
The upper
display shows the probe slope.
See figure
5.
Press the ENTER key.
The upper
display shows the % saturation offset.
See directions for setting this offset on
page 40.
See figure
6.
Press the ENTER key.
The upper
display shows the 100 % saturation mV
value.
See figure
7.
Press the ENTER key.
The upper display
shows the 0 % saturation mV value.
See figure
8.
Press the ENTER key
to move back to
subgroup P4.0. If you want to return to
measurement mode, press CAL/MEAS.
% saturation mode—
P4.0: Electrode properties
These “view only” parameters show the electrode properties for diagnostic purposes:
probe slope:
shows the calibration slope of the probe. This mode displays slope
from 0.5 to 1.999 % of slope (1.0 = 100%).
% saturation offset
: shows the value of the % saturation offset entered in parameter
P2.0 (see page 40 for instructions).
100% saturation mV value:
shows the
sensor’s mV output corresponding to
100% saturation.
0% saturation mV value:
shows the
sensor’s millivolt output corresponding
to 0% saturation.