P2.2: Electrode slope
NOTE: this parameter is available in
pH measurement mode only.
From measurement mode:
1.
If necessary, press the MODE key to
select pH measurement mode. Press
the SETUP key
to enter SETUP mode.
2.
Press the
▲
and
▼
keys
to scroll
through subgroups until you view
parameter P2.0.
See figure
3.
Press the ENTER key
twice to select
parameter P2.2.
4.
The display shows electrode slope
in percentage of the range you are
currently measuring. Default setting
is 100.0.
See figure
5.
Press the ENTER key
to return to the
subgroup menu. Press the CAL/MEAS
key to return to measurement mode.