23
If you need to set a very large K-factor (larger than 9999.9),
with a standard unit, you will have to follow the instructions for
getting an output pulse every 1/10th unit. That is, multiply the
meter’s K-factor by 0.1, and then enter the result on the switches.
Remember that you will get an output pulse every 1/10th unit
instead of every whole unit.
If you need to set a very large K-factor (greater than 9999.9), and
still need an output pulse for every whole unit, please contact
the factory. It is possible to supply, on special order, units with
a K-factor range of 00000 to 9999-9 (i.e., no decimal place).
Check cable-entry seals periodically. Tighten and/or apply seal-
ant if needed. This is especially important in environments con-
taining heavy concentrations of dust, oil mist or other residue.
Check all wiring connections occasionally for oxidation or cor-
rosion. Clean and re-seat if such conditions are noted.
If necessary, check and re-seat any connections that may have
been subjected to strain (during rework or construction, for
example).
MAINTENANCE