30
Control Tests
4
Troubleshooting Checks
Action
1. Were the test strips or
control solutions
expired?
Discard the test strips or control solution if either is past
the use by date. If the control solution was opened more
than 3 months ago, discard it. Repeat the control test
with an unexpired test strip and an unexpired control
solution.
2. Did you wipe the tip of the control
solution bottle before use?
Wipe the tip of the bottle with a tissue. Repeat the
control test with a new test strip and a fresh drop of
control solution.
3. Were the caps on the test strip container
and the control solution bottle always
closed tightly?
Replace the test strips or control solution if you think
either was uncapped for some time. Repeat the control
test.
4. Was the test strip used immediately after
it was removed from the test strip
container?
Repeat the control test with a new test strip and a fresh
drop of control solution.
5. Were the test strips and control solutions
stored in a cool, dry place?
Repeat the control test with a properly stored test strip
or control solution.
6. Did you follow the directions?
Read Chapter 4, Control Tests, and repeat the control
test.
7. Did you choose the correct control
solution level, either 1 or 2, when you
performed the control test?
If you chose the wrong control solution level, you can
still compare the control result to the range printed on
the test strip container.
8. Are you still unsure of the problem?
Contact Roche.
Understanding out-of-Range Control Results
w
WARNING
The control ranges are printed on the test strip container label. If the control result is out of range,
check this list to help solve the problem.
51820_07054190001_01_EN.indb 30
9/4/13 9:04 AM
Summary of Contents for Perfoma
Page 2: ...51820_07054190001_01_CVR indb 2 9 4 13 9 01 AM ...
Page 4: ...51820_07054190001_01_EN indb 2 9 4 13 9 03 AM ...
Page 38: ...36 Maintenance and Troubleshooting 5 51820_07054190001_01_EN indb 36 9 4 13 9 04 AM ...
Page 44: ...42 Technical Information 6 51820_07054190001_01_EN indb 42 9 4 13 9 04 AM ...
Page 46: ...44 Notes 51820_07054190001_01_EN indb 44 9 4 13 9 04 AM ...
Page 47: ...45 Notes 51820_07054190001_01_EN indb 45 9 4 13 9 04 AM ...
Page 48: ...46 Notes 51820_07054190001_01_EN indb 46 9 4 13 9 04 AM ...
Page 49: ...47 Notes 51820_07054190001_01_EN indb 47 9 4 13 9 04 AM ...
Page 50: ...48 Notes 51820_07054190001_01_EN indb 48 9 4 13 9 04 AM ...
Page 52: ...51820_07054190001_01_DE indb 2 9 4 13 9 06 AM ...
Page 86: ...36 Wartung und Fehlerbeseitigung 5 51820_07054190001_01_DE indb 36 9 4 13 9 06 AM ...
Page 92: ...42 Technische Informationen 6 51820_07054190001_01_DE indb 42 9 4 13 9 06 AM ...
Page 94: ...44 Notizen 51820_07054190001_01_DE indb 44 9 4 13 9 06 AM ...
Page 95: ...45 Notizen 51820_07054190001_01_DE indb 45 9 4 13 9 06 AM ...
Page 96: ...46 Notizen 51820_07054190001_01_DE indb 46 9 4 13 9 06 AM ...
Page 97: ...47 Notizen 51820_07054190001_01_DE indb 47 9 4 13 9 06 AM ...
Page 98: ...48 Notizen 51820_07054190001_01_DE indb 48 9 4 13 9 06 AM ...