Control Tests
4
8
Press and release to set the control level in the
meter.
oK
and the control result alternate on the display if
the control result is in range.
err
and the control result alternate on the display if
the control result is not in range.
Remove and discard the used test strip.
After a successful test, the meter turns off 5 seconds
after the test strip is removed.
7
The control result, the bottle symbol, and a flashing
L
appear on the display. Do not remove the test strip
yet.
Press to mark the control result as a Level 1.
Press a second time to mark the control result as
a Level 2.
or
(for example)
or
(for example)
51820_07054190001_01_EN.indb 29
9/4/13 9:04 AM
Summary of Contents for Perfoma
Page 2: ...51820_07054190001_01_CVR indb 2 9 4 13 9 01 AM ...
Page 4: ...51820_07054190001_01_EN indb 2 9 4 13 9 03 AM ...
Page 38: ...36 Maintenance and Troubleshooting 5 51820_07054190001_01_EN indb 36 9 4 13 9 04 AM ...
Page 44: ...42 Technical Information 6 51820_07054190001_01_EN indb 42 9 4 13 9 04 AM ...
Page 46: ...44 Notes 51820_07054190001_01_EN indb 44 9 4 13 9 04 AM ...
Page 47: ...45 Notes 51820_07054190001_01_EN indb 45 9 4 13 9 04 AM ...
Page 48: ...46 Notes 51820_07054190001_01_EN indb 46 9 4 13 9 04 AM ...
Page 49: ...47 Notes 51820_07054190001_01_EN indb 47 9 4 13 9 04 AM ...
Page 50: ...48 Notes 51820_07054190001_01_EN indb 48 9 4 13 9 04 AM ...
Page 52: ...51820_07054190001_01_DE indb 2 9 4 13 9 06 AM ...
Page 86: ...36 Wartung und Fehlerbeseitigung 5 51820_07054190001_01_DE indb 36 9 4 13 9 06 AM ...
Page 92: ...42 Technische Informationen 6 51820_07054190001_01_DE indb 42 9 4 13 9 06 AM ...
Page 94: ...44 Notizen 51820_07054190001_01_DE indb 44 9 4 13 9 06 AM ...
Page 95: ...45 Notizen 51820_07054190001_01_DE indb 45 9 4 13 9 06 AM ...
Page 96: ...46 Notizen 51820_07054190001_01_DE indb 46 9 4 13 9 06 AM ...
Page 97: ...47 Notizen 51820_07054190001_01_DE indb 47 9 4 13 9 06 AM ...
Page 98: ...48 Notizen 51820_07054190001_01_DE indb 48 9 4 13 9 06 AM ...