![background image](http://html.mh-extra.com/html/accu-chek/compact-plus/compact-plus_user-manual_455921190.webp)
190
Symbols, error messages and troubleshooting
Farbe: P 249 – Prüfmittelnummer: 203
Error message
Meaning
The batteries (regular or rechargeable)
are empty or a meter error has
occurred.
Turn the meter off and then on again
by pressing the Test button. If the error
message appears again, the batteries
may be empty. Insert new batteries
(new or recharged). If the error
message is again repeated when the
meter turns on, please contact the
Accu-Chek Customer Care Line.
The meter was exposed to a strong
electromagnetic field.
Press the Test button to turn the meter
off. Move somewhere else or turn off
the source of the field.
You bent or moved the test strip during
the test either while you were drawing
up blood or control solution with it or
during the test.