148
Changing the batteries
Farbe: P 249 – Prüfmittelnummer: 203
the test results saved in memory will be numbered consecutively
so you can still recognize the order in which they were saved
(see Chapter 5.2).
If the year, time and date have been lost, the meter rotates the
drum in order to re-read the bar code.
i
You can also use rechargeable NiMH batteries. If you do, please
note the following:
The number of tests you can perform with recharged
batteries is considerably less than with ordinary batteries.
To be ready to test at any time, you need to keep two
rechargeable batteries in reserve which you can charge while
the first pair are in the meter.
Recharged batteries lose their charge during storage.
When the battery symbol first appears, you can perform
significantly fewer than 50 further tests. With old and
exhausted rechargeable batteries you may not be able to
perform even one more test.
man_06919570001_01_EnCan.indd 148
18.01.2013 18:38:52