99
Appendix
Fig. 5: Subsequent boluses
hyper
hypo
Target Value
Target Range
bG level
Bolus
Current
bG values
Where the measured value is greater than the currently allowed bG value, the
newly calculated bolus (light-blue) only considers the difference between the
current bG value (dotted line) and the currently allowed bG value (the height of
the blue area). During the Acting Time of the first correction bolus (the width
of the blue area), only the currently allowed bG value is used to calculate the
second bolus.
Summary of Contents for Aviva
Page 1: ...Training Handbook mmol L Accu Chek Aviva Combo blood glucose monitoring system...
Page 3: ......
Page 13: ...13 Getting bolus advice from your meter...
Page 25: ...25 Notes...
Page 38: ...38...
Page 57: ...57 Managing your data...
Page 58: ......
Page 76: ......