98
Fig. 4: Corrected high bG
hyper
Target Value
Acting Time
Offset Time
Target Range
hypo
Corrected bG
bG level
Bolus
Current
bG values
Bolus
not required
Currently allowed
bG value: corrected bG
Time
The above diagram shows an example of the effect of this rule. The first correc-
tion bolus at 12 o’clock remains active during the Acting Time (the width
of the blue area). If a measured value at 2 o’clock falls within the currently
allowed bG value (height of the blue area), no new correction bolus is calcu-
lated.