140
Cleaning the meter and lancing device
Farbe: P 249 – Prüfmittelnummer: 203
141
Cleaning the meter and lancing device
8
8
Cleaning the meter and lancing device
8.1 Cleaning the meter
Your meter uses an optical measuring method that relies heavily on all
of its components being clean. Owing to the way the test strip works,
the meter does not normally come into direct contact with blood or con-
trol solution. Regular cleaning is therefore largely unnecessary. In the
event of the meter becoming soiled through improper use, cleaning it
may become necessary.
Wipe away any soiling from the housing or display with a cloth lightly
moistened with cold water or 70 % ethanol.
Clean the test strip guide and measuring window thoroughly if
O
the test strip guide has become visibly soiled,
O
error message E-5 is displayed (see Chapter 12.2) and the soiled
measuring window is the cause.
Do this as follows:
P
If the meter is on, press the Test button to turn it off.
P
Wait for the motor to stop.
P
Wipe the test strip guide with a cotton swab lightly moistened with
cold water or 70 % ethanol.
P
Place your thumb into the recess below the drum compartment
cover and fold the drum compartment cover upwards.