Syst e m re qu ir e m en t s
N or m a l ope r a t in g e n viron m en t :
Tem perat ure + 10 t o + 40° C.
Relat ive hum idit y 20-80% non- condensing.
Below 2500 m above sea level.
Sh or t te r m st or a ge / t ra n spor t a t ion
e n vir on m e n t ( u p t o 9 6 h ou r s) :
I n packaging- box/ shipper- box.
Maxim um - 30 t o + 70° C.
Maxim um 93% RH condensing, at + 40° C.
Minim um am bient pressure
59,7 kPa ( 4.267
m above sea level) .
N or m a l st or a ge e n vir on m e n t :
Tem perat ure + 10 t o + 40° C.
Relat ive hum idit y 20-80% non- condensing.
I n gr e ss Prot e ct ion ( I P) r a t in gs:
CADScor
®
Sensor I P44
CADScor
®
Docking st at ion I P22
Power adapt ors I P20
Ope r a t in g noise en vir on m en t :
Ext ernal noise level < 65dB SPL.
Ext ernal elect rical noise level < 65dB SPL.
Syst e m a lt it u de cor r ect ion
Up t o 2500 m above sea level
Syst e m pow er sour ce
Only t he supplied power adapt or m ust be
used, ident ified by a m anufact urer nam e and
t ype num ber:
XP- Power ACM06US05 ( - XZ1505)
Rat ed input volt age: 100-240VAC, 50/ 60 Hz
Rat ed out put volt age: 5VDC
Max. power: 6.0 W
Never use ot her power adapt or t han
supplied wit h t he CADScor
®
Syst em , due t o
risk of elect rical shock.