7.14
Re- set t ing sensor
The CADScor
®
Sensor operat ing syst em can
be reset in t he event of unresponsive
soft ware failure. The re-set t ing will force
t he sensor t o reload sensor - soft ware and
also re- qualify t he sensor.
Procedure:
Press and hold t he push but t on for 8
seconds ( beyond t he 4 seconds single
beep, unt il a double beep signal feedback
is given) t o power off t he sensor.
The sensor will aut om at ically t urn on again
while in t he powered Docking st at ion or by
but t on press.
The sensor will t hen reload operat ing
soft ware and qualify aut om at ically.
The sensor will display t he st art -up display
( REF: 101) followed by t he " last recording"
display ( REF: 103) when t he qualificat ion
has finished, and sensor is docked.
3.2