Er r or m essag e
D isp la y
r e fe r e n ce
Possib le ca use
Corr e ct ion a ct ion
# 20
The LED indicat or on
Dock ing st at ion is
off, or const ant
y ellow when Sensor
is docked.
The power adapt or has
been turned off at m ains.
The power j ack has been
disconnected.
Plug assem bly not fully
inserted int o adapt or.
The power adapt or is
fault y, or elect rical supply
is m issing.
The Dock ing stat ion
contacts interface is not
recognizing sensor.
Turn on power adapt or out let.
Apply power j ack supply t o t he dock ing
st at ion.
Est ablish full insert ion of plug int o power
adaptor by pressing firm ly .
Replace power supply unit if suspected
m alfunct ioning.
Restore elect rical supply.
Try rest art procedure ( 7.14) .
Cont act local dist ribut or/ Acarix for
available opt ions if problem persist s.
# 21
The display is black .
The sensor is in off m ode.
The sensor bat tery is
discharged.
See also # 3.
Press push butt on t o swit ch on sensor.
Charge t he sensor bat tery in t he Dock ing
st at ion.
For deep discharged bat teries recharging
m ay t ake 1- 2 day s.
I t is recom m ended t o always k eep power
on sy stem , for ease of use and batt ery
charging.
# 22
Display cracked
The display is brok en.
Cont act local dist ribut or/ Acarix for
available opt ions.
# 23
Sensor unresponsiv e
Soft ware error.
Sensor needs rest art. See sect ion 7.14