Er r or m essag e
D isp la y
r e fe r e n ce
Possib le ca use
Corr e ct ion a ct ion
# 16
Am bient m icrophone
error
REF: 216
The test signal received by
t he am bient m icrophone
was lower t han
ant icipated.
Error was t riggered by
liquid exposure to sensor.
Take care not t o cover sensor or am bient
m icrophone during recording.
Ensure CADScor
®
Sensor body is oriented
along sternum t owards chin.
Allow sensor to dry > 24 hours in Dock ing
st at ion.
I f persistent failure cont act local
dist ribut or/ Acarix serv ice.
# 17
Self- t est failed
REF: 315
A t echnical problem inside
t he CADScor
®
Sy stem has
occurred.
Cont act local dist ribut or/ Acarix serv ice.
# 18
Unexpected issue
det ected. Rest art
required.
REF: 317
An int ernal error has
occurred. Sensor needs t o
be restarted.
The USB connect ion
bet ween Dock ing stat ion
and Sensor failed.
Place sensor in dock ing st at ion and press
“ EXI T” .
Wait for restart t o finish.
I nspect charging pins in Dock ing st at ion
for free up and down m ov em ent . Exercise
pin gent ly t o free up stuck pin.
Clean charging pads on sensor using
cot t on swabs wit h 70% et hanol.