background image

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Instructions 

F16LSC, F16LSCN

Calibration Preparation

Gauges are calibrated at the factory using equipment traceable to NIST. 

There is no need to calibrate the gauge before putting it into service. 
Calibration should only be performed by qualified individuals using 

appropriate calibration standards and procedures.
Gauges can be returned to factory for certified recalibration and 

repairs. NIST traceability is available.
Calibration intervals depend on your quality control program 

requirements and as-found data. Many customers calibrate their 

equipment annually.
The calibration equipment should be at least four times more accu-

rate than the gauge being calibrated.
The calibration system must be able to generate and measure pres-

sure and/or vacuum over the full range of the gauge. 
A vacuum pump able to produce a vacuum of 100 microns (0.1 

torr or 100 millitorr) or lower is required for vacuum and absolute 

gauges.
Use a stable DC power supply and an accurate mA meter to cali-

brate loop powered transmitters.
Allow the gauge to equalize to normal room temperature for at least 

20 minutes before calibration.

The gauge must be powered by a 10-32 VDC power supply during 

calibration.

Enter Calibration Pass Code

User-Defined Pass Code Configuration

SEL TEST

t

s

0.0

 

*

*

<and>

then

*

The factory default 3510 pass code may be changed to a differ-

ent value.

Configuration Pass Code

With the unit on, press and hold the 

s

 and TEST buttons and 

then press the SEL button. Release all buttons when the display 

indicates 

CFG

.

Calibration Pass Code

With the unit on, press and hold the 

t

 and TEST buttons and 

then press the SEL button. Release all buttons when the display 

indicates 

CAL

.

Test Function Pass Code (if enabled)

With the unit on,, press and hold the TEST button and press the 

SEL button. Release both buttons when the upper display indicates 

– – – –

 .

Change Pass Code Mode

Before the unit enters the view or change pass code mode, the 

display initially indicates _ _ _ _ with the first underscore blinking, 

and with 

CFGPC

CALPC

, or 

TSTPC 

on the lower display.

The gauge will revert to normal operation if no buttons are operated 

for approximately 15 seconds. Press and release the SEL button 

without entering any pass code characters to exit.
Enter access code 1220:
Use the 

s

 and 

t

 buttons to set the left-most digit to 1.

Press and release the SEL button to index to the next position. The 

1 will remain, and the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the SEL button to index to the next position. 1 2 

will remain, and the third position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the SEL button to index to the next position. 1 2 

2 will remain, and the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the SEL button to proceed. 
Note: If an incorrect access code was entered, the gauge will return 

to the start of the access code entry sequence.

Change Pass Code

Once the access code has been entered correctly, the display 

will indicate the existing user-defined pass code with 

CFGPC

CALPC

, or 

TSTPC 

on the lower display.

Press the 

s

 or 

t

 button to select the first character of the new 

pass code.
When the correct first character is being displayed, press and 

release the SEL button to proceed to the next pass code character.
Repeat above until the entire pass code is complete.
To exit the User-Defined Pass Code change mode, press and hold 

the SEL button until the gauge restarts. 

With the gauge powered up, press and hold the 

t

 

and TEST buttons.
Then press the SEL button to enter the Calibration 

Mode. 

Release all buttons when the display indicates 

CAL

. The firmware version is also displayed.

SEL TEST

t

s

CAL

 V 04E 

The display prompts for entry of the calibration 

pass code (

CALPC

), with the first underscore 

blinking. To enter the 4 digit pass code, see the 

Enter Configuration Pass Code section. 

Note that 

it is possible to change the default 3510 value.

____

 CALPC

Calibration Mode

The gauge enters and remains in the Calibration Mode until 

restarted manually or power is removed. 
The display will then indicate the currently applied pressure in the 

engineering units selected in gauge configuration.

Button Operation

Each time the 

s

 or 

t

 button is pressed and released quickly, a 

small change is made to the digitized pressure signal. 
It may take more than one of these small changes to result in a 

single digit change on the display. 
To make larger changes, press and hold the appropriate button. 

After about one second, the display will begin to change continu-

ously. Release the button to stop. 
Then make fine adjustments by pressing and quickly releasing the 

s

 or 

t

 buttons as previously described.

The SEL button is used to select 

LCAL

 (4 mA calibration), 

HCAL

 

(20 mA calibration), or 

CAL

 (pressure calibration).

If the SEL button is depressed for longer than 2 seconds, the 

display will change to indicate – – – – , and the gauge will exit the 

Calibration Mode when the button is released.

Loop Current Calibration

Loop current calibration coordinates the loop current to the display 

indication, and is performed independently of applied pressure. It 

requires a direct physical measurement of the loop current. See 

wiring diagrams on page 2.

4 mA Loop Current

Press and release the SEL button until the display 

briefly indicates 

LCAL

.

Note: If the SEL button is depressed for lon-

ger than 2 seconds, the gauge will exit the 

Calibration Mode.

SEL TEST

t

s

LCAL

 V 04E 

*

SEL TEST

t

s

0.00

 cal 

The upper display will indicate the pre-configured 

pressure corresponding to a 4 mA loop current. 
The lower display segments will alternate 

between 

CAL

 and 

4 MA

.

Use the 

s

 or 

t

 buttons to adjust the actual loop 

current to 4 mA.

 4 mA

or

*

*

Power (+) 

Power (–)

0.00

mA meter 

4.000

20 mA Loop Current

Press and release the SEL button until the display 

briefly indicates 

HCAL

.

Note: If the SEL button is depressed for lon-

ger than 2 seconds, the gauge will exit the 

Calibration Mode.

SEL TEST

t

s

HCAL

 V 04E 

*

SEL TEST

t

s

The upper display will indicate the pre-configured 

pressure corresponding to a 20 mA loop current. 
The lower display segments will alternate 

between 

CAL

 and 

20 MA

.

Use the 

s

 or 

t

 buttons to adjust the actual loop 

current to 20 mA.

i00.0

 cal 

 20 mA

or

*

*

Power (+) 

Power (–)

i00.0

mA meter 

20.00

Pressure Calibration

The pressure calibration procedure simultaneously adjusts both the 

display indication and the loop current to correspond to the actual 

applied pressure.

Press and release the SEL button until the display 

briefly indicates 

CAL

.

SEL TEST

t

s

CAL

 V 04E 

*

SEL TEST

t

s

0.00

 cal 

Zero Calibration

Apply zero pressure.
The lower display segments will alternate 

between 

CAL

 and 

ZERO

.

Use the 

s

 or 

t

 buttons to adjust the display 

to zero.

 zero

or

*

*

Span Calibration

Apply full scale pressure.
The lower display segments will alternate 

between 

CAL

 and 

+SPAN

.

Use the 

s

 or 

t

 buttons to adjust the display to 

match the calibrator's pressure reading.

SEL TEST

t

s

0.00

 cal 

+span

or

*

*

Mid Point Calibration

Apply 50% of full scale pressure.
The lower display segments will alternate 

between 

CAL

 and 

+MID

.

Use the 

s

 or 

t

 buttons to adjust the display to 

match the calibrator's pressure reading.

SEL TEST

t

s

0.00

 cal 

+MID

or

*

*

Negative Span Calibration

This applies to bipolar and compound ranges only.

Apply full scale negative pressure (full vacuum).
The lower display segments will alternate 

between 

CAL

 and 

-SPAN

.

Use the 

s

 or 

t

 buttons to adjust the display to 

match the calibrator's pressure reading.

SEL TEST

t

s

0.00

 cal 

-SPAN

or

*

*

Negative Mid Point Calibration

This applies only to bipolar ranges that use the 

±15 psig sensor.

Apply 50% full scale negative pressure (50% 

vacuum).
The lower display segments will alternate 

between 

CAL

 and 

–MID

.

Use the 

s

 or 

t

 buttons to adjust the display to 

match the calibrator's pressure reading.

SEL TEST

t

s

0.00

 cal 

-mid

or

*

*

To store the calibration parameters and exit the 

Calibration Mode, press and hold the SEL button 

until the display indicates 

– – – –

 . The gauge 

will restart.
Verify the pressure indications at 0%, 25%, 50%, 

75% and 100% of full range of the gauge.

SEL TEST

t

s

0.00

 cal 

-mid

press and hold 

until gauge restarts

*

Loop Current Calibration—continued

continued 

next column

To store the calibration parameters and exit the 

Calibration Mode, press and hold the SEL button 

until the display indicates 

– – – –

 . The gauge 

will restart.
Verify output at 0%, 25%, 50%, 75% and 100% 

of full range of the gauge.

Summary of Contents for Cecomp F16LSC

Page 1: ...of model number HA High accuracy 0 1 FS 1 LS See table at left for availability MC Metal front cover instead of plastic n a NEMA 4X PM Panel mount 4 1 x 4 1 n a NEMA 4X CC Moisture resistant circuit board conformal coating CD Calibration data 5 test points and date NC NIST traceability documentation 5 points and date SCR14SS Filter screen fitting keeps debris out of gauge sensor Use for food vacuu...

Page 2: ...15psi 4 mA 12 mA 20 mA Compound 30inHg 100psi 4 mA 5 5 mA 20 mA Compound 30inHg 200psi 4 mA 4 8 mA 20 mA Absolute reference 4 mA 4 mA 20 mA Bipolar 4 mA 12 mA 20 mA cecomp com 1220 American Way Libertyville IL 60048 800 942 0315 CECOMP Div of ABSOLUTE PROCESS INSTRUMENTS 1 65 2 88 NEMA 4X 3 0 2 0 Turn at hex fitting only Turn at hex fitting only 3 ft long 2 conductor 22 AWG cable 3 ft long 2 condu...

Page 3: ...will be off The pressure will display for 15 seconds then the gauge will shut off SEL TEST t s SEL TEST t s on hold SEL TEST t s OFF SEL TEST t s i00 0 psig After initialization the gauge will indicate the applied pressure for 15 seconds or The gauge must be powered by a loop supply during configuration Release all buttons when the display indicates CFG The gauge firmware version is also dis playe...

Page 4: ...l restarted manually or power is removed The display will then indicate the currently applied pressure in the engineering units selected in gauge configuration Button Operation Each time the s or t button is pressed and released quickly a small change is made to the digitized pressure signal It may take more than one of these small changes to result in a single digit change on the display To make ...

Reviews: