Zero/Tare Mode

Zero/Tare applies to gauge reference models only. Absolute refer-

ence gauges do not use the zero feature since they read atmo-

spheric pressure under normal conditions.
This feature can be enabled or disabled in Gauge Configuration.
The gauge must be in the Normal Mode with the gauge port 

exposed to normal atmospheric pressure.
While in the Zero Tare mode, the gauge will not respond to any 

changes in pressure. The loop current will maintain its last value.

Gauge Configuration

Press and hold both the 

s

 and 

buttons.

Then press the SEL button.

hold

*

*

then 

press

Release all buttons when 

oooo

 is displayed.

SEL TEST

t

s

oooo

 psig

The display will indicate a new zero tare value 

with 

Z OFF

 (zero offset) on the lower display. 

To exit with no changes, press SEL.
If not within approximately 3% 

full scale of zero, 

Err0 

will be 

displayed and no changes are 

allowed. Pressure must be removed from the 

gauge, or it must be recalibrated.

SEL TEST

t

s

0. I 

Z OFF

or

SEL TEST

t

s

0. 0 

Z off

*

To remove the existing zero tare value, press and 

release the 

button. The display changes to 

zero. The Normal Mode may indicate a non-zero 

value since zero correction has been removed.

To restore the newly calculated zero tare value, 

press and release the 

button.

SEL TEST

t

s

0. I 

Z off

*

SEL TEST

t

s

0. 0 

Z off

exit

*

To exit the Zero Tare mode and return to the 

Normal Mode, press and release the SEL button.

E r r

0

If excessive vacuum is applied to a pressure-

only gauge, the display will indicate an out-of-

range indication of 

–Err

. Applying vacuum to 

a pressure-only gauge can damage the sensor.
If excessive vacuum is applied to a vacuum-

pressure gauge, the display will indicate an out-

of-range indication of 

– I – – –

 or 

–I

.–.–.– will 

be displayed depending on model.
If 112.5% over-range pressure is applied, an 

out-of-range indication of 1 – – – or 1.–.–.– will 

be displayed depending on model.

Out-of-Range Indications

-E

r r 

psig

-I- - - - 

psig

I.-.-.-.- 

psig

I- - - - 

psig

exit

Test Function

The Test Mode will allow setup and testing 

of the current loop without having to alter the 

system pressure. 
From the Normal Mode with applied pressure 

being displayed, press and hold the TEST button.

hold

SEL TEST

t

s

50.0

 psig

*

SEL TEST

t

s

i00.0

 test

While holding the TEST button, press the SEL 

button.

SEL TEST

t

s

i00.0

 TEST

both

SEL TEST

t

s

50.0

 psig

*

*

- - - -

When the display indicates – – – – , release 

both buttons.
If a pass code is required, the upper display will 

indicate _ _ _ _ with the left-most position blink-

ing, and the lower section will indicate 

TSTPC

  

(Test Pass Code). See the "Enter Configuration 

Pass code" section of this manual to enter the 

pass code and continue.
Next, the upper display will indicate the applied 

pressure and the units display will be blinking.

While the TEST button is pressed, the display and 

loop current are switched, independent of the 

actual pressure, to a level determined by the test 

setting. 

TEST

 is indicated on the lower display. 

Note: The gauge will not respond to changes in 

applied pressure while the TEST button is held.

Pressing the 

s

 or 

buttons while pressing the 

TEST button will raise or lower the test value. 
When the TEST button is released, the setting is 

stored in non-volatile memory and the gauge will 

operate normally.
Press SEL to exit the Test Mode and return to 

the Normal Mode.

50.0

 psig

____

TSTPC

or

*

or

*

*

hold

*

SEL TEST

t

s

*

SEL TEST

t

s

0. I 

 psig

Backup Power Mode Operation

The power supply for the Backup Power mode takes approximately 

two hours of operation in the Normal Mode to fully charge the 

backup power supply from the loop supply.

To power up the gauge, press and release the 

SEL button. 

To power-up the gauge for a longer period of time 

and override the auto shutoff, press and hold the 

SEL button until the display indicates 

On

The gauge will indicate the pressure until the 

backup power is depleted. A fully charged 

backup supply will last up to 40 minutes.
Press SEL to manually power 

OFF

 the gauge.

The gauge will return to Normal Mode when loop 

power is restored.

SEL TEST

t

s

i00.0

 psig

If loop power is lost, the gauge will automatically 

switch to the Backup Power mode. 

The low power indicator will flash, and the Loop-

Tracker LED will be off. The pressure will display 

for 15 seconds, then the gauge will shut off.

SEL TEST

t

s

*

SEL TEST

t

s

o

n

*

hold

SEL TEST

t

s

OFF 

*

SEL TEST

t

s

i00.0

 psig

After initialization, the gauge will indicate the 

applied pressure for 15 seconds.

or

The gauge must be powered by a loop supply during configuration.

Release all buttons when the display indicates 

CFG

. The gauge firmware version is also dis-

played.

SEL TEST

t

s

CFG

 V 04E

The display prompts for entry of the configura-

tion pass code (

CFGPC

), with the first under-

score blinking. To enter the 4 digit pass code, 

see the Enter Configuration Pass Code section. 

____

 CFGPC

The first position will be blinking.

The 3 will remain, and the second position will 

be blinking. 

Enter Configuration Pass Code

SEL TEST

t

s

3___

 

*

*

When a pass code is required, the upper display will indicate _ _ _ _ , 

and the lower display will indicate 

CFGPC

CALPC

, or 

TSTPC

 

depending on the feature being accessed. 
To cancel, press the SEL button without entering any numbers. If no 

buttons are pressed, the gauge will revert to normal operation after 

15 seconds. To proceed, enter the user-defined pass code. 3510 is 

the factory default, but it is user-modifiable.

*

3___

 

SEL TEST

t

s

35__

 

*

*

*

SEL TEST

t

s

35 I_

 

*

*

*

35__

 

Note: If an incorrect pass code is entered, the 

gauge will return to the start of the pass code 

entry sequence.

Use the 

s

 or 

t

 buttons to select 0.

Press and release the SEL button to proceed with 

configuration procedures. 

Use the 

s

 or 

t

 buttons to select 1.

Press and release the SEL button to index to the 

next position. 

____

 

Use the 

s

 or 

t

 buttons to set the left-most 

digit to 3.
Press and release the SEL button to index to the 

next position.

Use the 

s

 or 

t

 buttons to select 5.

Press and release the SEL button to index to the 

next position. 

3 5 will remain, and the third position will be 

blinking. 

3 5 1 will remain, and the fourth position will 

be blinking. 

35 I_

 

SEL TEST

t

s

35 I0

 

*

*

*

then

then

then

then

or

or

or

or

Press and hold the TEST and 

s

 buttons, then 

press the SEL button.

SEL TEST

t

s

i00.0

 

*

*

and

then

*

*

Factory/User Configuration 

Upon successful pass code entry, the upper 

display will be blank, and the lower will display 

USER

. Use the 

t

 or 

s

 buttons to select 

USER

 or 

FCTRY

With 

USER

 selected, the gauge configuration 

can be modified as described below. 
To reset the gauge to the factory configuration, 

press SEL while 

FCTRY

 is displayed. The gauge 

will restart with the factory configuration restored.

 USER

SEL TEST

t

s

 FCTRY

* *

or

or

then

*

Gauge Type Configuration 

This applies to gauge reference vacuum/pres-

sure models only.

Use the 

t

 or 

buttons to select:

-/+EU

 

Vacuum is indicated as a negative 

number in the selected engineering units

CMPND

  Vacuum is displayed as negative 

InHg and pressure in PSIG.
Press and release the SEL button to move to the 

next parameter.

+/- EU

SEL TEST

t

s

CMPND

* *

or

or

then

*

Engineering Unit Selection

This is skipped if 

CMPD

 was selected.

Use the 

t

 or 

buttons to select the engineer-

ing units available for the range of the gauge:
Press and release the SEL button to move to the 

next parameter.

PSIG

SEL TEST

t

s

* *

or

then

*

Zero Tare Enable/Disable

This is not used for absolute reference gauges.

Use the 

t

 or 

buttons to select:

ZTARE

 

Zero Tare function enabled.

NOZTR

 

Zero Tare function disabled.

Press and release the SEL button to move to the 

next parameter.

ZTARE

SEL TEST

t

s

NOZTR

* *

or

or

then

*

Test Function Pass Code Enable/Disable

Use the 

t

 or 

buttons to select:

TSTPC

  Pass code required for Test Mode.

NOTPC

  No pass code required for Test Mode.

Press and release the SEL button to move to the 

next parameter.

TSTPC

SEL TEST

t

s

NOTPC

* *

or

or

then

*

o.00

RNGLO

SEL TEST

t

s

* *

or

then

*

Range Lower Limit Adjust

The upper display will indicate the pressure 

value corresponding to 4 mA loop current. The 

lower section will display 

RNGLO

.

Use the 

t

 or 

buttons to display the desired 

pressure equal to a 4 mA output:
Press and release the SEL button to move to the 

next parameter.

o.00

RNGHI

SEL TEST

t

s

* *

or

then

*

Range Upper Limit Adjust

The upper display will indicate the pressure 

value corresponding to 20 mA loop current. The 

lower section will display 

RNGHI

.

Use the 

t

 or 

buttons to display the desired 

pressure equal to a 20 mA output:
Press and release the SEL button to save the 

user configuration and restart the gauge. 

The configuration parameters will not be saved if 

the procedure is interrupted before completion.

i00.0

 PSIG

SEL TEST

t

s

restart

normal operation

-I.-.-.-.- 

psig

Summary of Contents for Cecomp F16LSC

Page 1: ...of model number HA High accuracy 0 1 FS 1 LS See table at left for availability MC Metal front cover instead of plastic n a NEMA 4X PM Panel mount 4 1 x 4 1 n a NEMA 4X CC Moisture resistant circuit board conformal coating CD Calibration data 5 test points and date NC NIST traceability documentation 5 points and date SCR14SS Filter screen fitting keeps debris out of gauge sensor Use for food vacuu...

Page 2: ...15psi 4 mA 12 mA 20 mA Compound 30inHg 100psi 4 mA 5 5 mA 20 mA Compound 30inHg 200psi 4 mA 4 8 mA 20 mA Absolute reference 4 mA 4 mA 20 mA Bipolar 4 mA 12 mA 20 mA cecomp com 1220 American Way Libertyville IL 60048 800 942 0315 CECOMP Div of ABSOLUTE PROCESS INSTRUMENTS 1 65 2 88 NEMA 4X 3 0 2 0 Turn at hex fitting only Turn at hex fitting only 3 ft long 2 conductor 22 AWG cable 3 ft long 2 condu...

Page 3: ...will be off The pressure will display for 15 seconds then the gauge will shut off SEL TEST t s SEL TEST t s on hold SEL TEST t s OFF SEL TEST t s i00 0 psig After initialization the gauge will indicate the applied pressure for 15 seconds or The gauge must be powered by a loop supply during configuration Release all buttons when the display indicates CFG The gauge firmware version is also dis playe...

Page 4: ...l restarted manually or power is removed The display will then indicate the currently applied pressure in the engineering units selected in gauge configuration Button Operation Each time the s or t button is pressed and released quickly a small change is made to the digitized pressure signal It may take more than one of these small changes to result in a single digit change on the display To make ...

Reviews: