Symptom:
Green, Red or Amber indicator lights “OFF”, POWER
Switch is in the
“ON” position and the cooling fan is not running.
1.
Check if the power coming to the generator.
2.
Check the Power fuse located on exterior of the Ozone Generator and if
the fuse is “OK” check power cord for breaks in insulation or loose
connections or contact the Service Department
3.
If the fuse is not “OK” please contact Service Department
Symptom:
Green indicator
light is “ON”, Red or
Amber indicator
light are “ON”,
Internal Safety Shutoff;
1.
Please check the following;
a.
Presence of moisture inside the tubing and the generator. If
moisture is present, dry it out by running oxygen though the
generator without power (that could take from a few hours to
several days). Do not power the generator unless absolutely sure
that it is dry. Running the generator with moisture inside will cause
serious damage to the internal components and costly repair.
b.
Red or Amber light can be “ON” also in case of Remote “Off”
condition
c.
Oxygen pressure and adjust the pressure according to the
specifications.
2.
If the above conditions are not present, please turn the Ozone Production
Potentiometer on the Control Panel ½ of a turn down and switch the
Power S
witch “ON” again. If the Red indicator light is still “ON” –
Please
contact our Service Department
If the above conditions are no longer present, please turn the power to “OFF”
position, turn the potentiometer to 0 position, wait 10 seconds and turn the
power switch “ON”. Adjust the potentiometer according to the instructions.
Should the problem persist, please contact our service department.