BiWaze Cough - User Reference Manual
60
•
Inhale Pressure not to exceed 85 cmH2O for 1 minute
•
Exhale Pressure not to exceed -75 cmH2O for 5 secs
•
Duration of inhale phase in Auto Mode within ± (10% of the s 0.5
seconds)
•
Duration of exhale phase in Auto Mode within ± (10% of the s 0.5
seconds)
•
All breath phases with times > 0 occurring in proper order in Auto Mode