6.6.3
Oscillation feature
BiWaze Cough device provides oscillation feature during all phases of a breath
cycle. The user can enable/disable the oscillations with toggle button on the
main screen.
The user can also change the Frequency of oscillations in 5-20 Hz Range
amplitude in levels 1-5.
6.6.4
Auto Mode – Advanced Cough Cycle Customization
BiWaze Cough Therapy device allows advanced customization of Cough Cycles
for advanced users. Example: to program last n cycles with only positive pressure
to increase patient comfort. This setting is available only in Unlocked Operation
Mode.
Once user has programmed the Auto mode therapy settings from main screen
including number of Cough breath cycles. User can initiate advanced Cough
Cycle customization by performing a “long touch” on the wave form area and
user is presented with following screen.
Vibration toggle
If this option is available in the Device Model. Allows you
to super impose low amplitude 20 Hz oscillations over
and above the set Oscillation frequency and amplitude.