![background image](http://html2.mh-extra.com/html/abbott/optis-1014933/optis-1014933_operation-and-maintenance-manual_5034488046.webp)
42
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
Table 7 (continued). Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is
used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.
Table 8. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is
used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the Transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
[V
1
= 3]
Recommended separation distance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
[E
1
= 3]
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where
P
is the maximum output power rating of the
Transmitter in watts (W) according to the Transmitter
manufacturer and
d
is the recommended separation
distance in meters (m)
Field strengths from fixed RF Transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
Immunity to Proximity
Fields from RF wireless
communications
equipment
IEC 60601-1-2
(Clause 8.10)
385-5785 MHz
9-28 V/m
Per IEC 60601-1-2 (Per Table 9 of Standard)
State: Released
Date: 2021.03.01 22:19 GMT
Effectivity: Upon Release
Summary of Contents for OPTIS 1014933
Page 2: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 4: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 8: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 49: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...