34
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
6.
Pull the adapter straight out of the DOC.
Note:
Ensure that the adapter does not rotate while it is being
removed, as this could damage the DOC or break the adapter.
7.
Remove the cap and plug from the replacement optical adapter.
Note:
Do not touch any of the optical connectors, as this may
damage them.
8.
Using fingers, align the new adapter with the DOC, ensuring that the key is
aligned with the slot in the optical carriage, and firmly press it into place. An
audible click is produced when the adapter seats.
Note:
If the carriage moves when you press on the adapter, click the
Exit button in the DOC Service section of the Service menu. When it
changes to Enter, click it again. When the optical carriage is locked
into position at the front of the DOC, firmly press the adapter into
place.
Figure 11. Alignment of Optical Adapter with Optical Carriage
9.
In the DOC Service section of the Service menu, click Exit. The DOC optical
carriage moves all the way to the back and restores all DOC functions to
their normal state.
10.
Click OK or Cancel to close the Setup dialog box.
11.
Return the system to normal operation.
1.
Engagement pin
2.
Catheter connection
port
1
2
State: Released
Date: 2021.03.01 22:19 GMT
Effectivity: Upon Release
Summary of Contents for OPTIS 1014933
Page 2: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 4: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 8: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...
Page 49: ...State Released Date 2021 03 01 22 19 GMT Effectivity Upon Release...