32
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
9.
In the DOC Service section of the Service menu, click Exit. The DOC optical
carriage moves all the way to the back and all DOC functions are returned
to their normal state.
10.
Click OK or Cancel to close the Setup dialog box.
11.
Open the sizing cap and place it on the end of the optical fiber connector
cleaner (figure below).
12.
Insert the cleaner into the optical connection in the catheter, ensuring that it
seats fully, and press until it clicks. The cleaning material in the optical fiber
connector cleaner is moved over the optical connection.
Figure 9. Inserting Cleaner into Catheter
13.
Remove the cleaner from the catheter.
14.
Reconnect the catheter to the DOC.
15.
Return the system to normal operation.
State: Released
Date: 2021.03.01 22:19 GMT
Effectivity: Upon Release