1
Call Customer Care Toll Free at 1-888-522-5226 www.AbbottDiabetesCare.com
EN
G
LI
SH
1
EN
G
LI
SH
The result will appear on the display screen and if the sound is on,
your meter will beep twice. To mark the result as a control solution
result, press and hold the “
c
” button for two seconds until the control
solution symbol ( ) appears at the top of the display screen.
If you have marked a blood glucose test as a control solution test by
mistake, you can unmark it by pressing and holding the “
c
” button
again for two seconds.
Compare your control solution test result to the
range printed on your test strip vial label. Your
control solution results should fall within the
range printed on the side of your test strip vial.
Out of range test results may be caused by:
• Expired or bad control solution.
• Expired or bad test strip.
• Error in performing test.
• Watered-down control solution.
• Code on test strip vial does not match code set in meter.
• Meter malfunction.
• Control solution test done outside 59° to 104° F.
To return the meter to clock mode, remove the test strip
.
The FreeStyle® Lancing Device
Dark Cap
(for finger testing)
Cocking Handle
Depth Settings Dial
Depth Indicator Window
Release Button
Clear Cap (for testing sites
other than the fingers)
Using the Lancing Device
1
Remove Cap
Snap off the cap on the lancing device at
an angle.
Insert Lancet
Insert a new FreeStyle Lancet firmly into
the white lancet holder cup. (Pushing the
lancet into the cup may cock the device;
this is okay.)
Twist Off the Rounded Top
Hold the lancet firmly in place with one
hand; use your other hand to twist off the
rounded top.