18
TROUBLESHOOTING GUIDE
SCREEN DISPLAY POSSIBLE CAUSE
ACTION TO TAKE
ATTN
PUMP ON HOLD
This means that hold time has
elapsed
Turn dial to RUN to restart
the pump
BATT
BATT EMPTY
This means the pump battery
is depleted and the pump has
stopped working
Plug the power cord into a mains outlet and
make sure the power is turned on. The display
will indicate CHRG PLEASE WAIT until pump
is ready to use
BATT
LOW BATTERY
This means the pump
battery is nearly empty and
approximately 20 minutes of
battery operation remain
Plug the power cord into a mains outlet and make
sure the power is turned on
CHRG
PLEASE WAIT
The battery charge is below
the minimum level to operate
and the pump cannot be used
until minimum charge level
is reached
No action to be taken. Pump is available for
use when the display indicates PUMP READY
TO USE
DIAL
NOT SET
Pump dial is between two dial
positions e.g. between RUN
and HOLD
Turn the dial to the desired position directly
aligning the two dots (on the dial and pump
casing). Continue setting up and operating pump
as normal
DOOR
OPEN
This means the door is open
while feeding or priming the
giving set
Turn dial to HOLD and close the set door. Turn
dial to RUN to restart pump or attempt to prime
the giving set again
DOSE
DONE
This shows the pump has
completed delivering the
programmed dose
If you have finished feeding:
turn dial to OFF/CHRG as the dose has been
delivered (VOL FED has reached your set dose).
Or, if there is more feed to deliver: turn dial to
CLEAR VOL (after 5 seconds, volume will be
cleared). Then turn dial to SET DOSE, and use
the Up and Down arrows to set the remaining
dose to be delivered. Turn dial to RUN to restart
the pump
ER##
SYSTEM FAIL
This indicates a
pump malfunction
You should remove the pump from the patient.
Turn the dial to OFF/CHRG and contact Abbott
If you have any problems setting up your FreeGo® pump and feed system,
please call Abbott Nutrition on Freephone 1800 22 11 66