![background image](http://html1.mh-extra.com/html/abbott/freego/freego_training-manual_451009017.webp)
Before cleaning make sure the pump is
off. Switch off mains power and unplug
from the wall socket and the pump
Wipe carefully and thoroughly with a
soft damp cloth and warm soapy water
EVERYDAY CARE OF YOUR PUMP
WARNING: Do not submerge the
pump or spray water or cleaning
solution into the pump.
For general disinfecting, use
10% solution of 5.25% sodium
hypochlorite (household bleach),
or 40% solution of ethyl alcohol,
or 10% solution of phosphoric acid.
When you have finished cleaning the
pump, it should be reconnected so that
the battery can continue charging
The pump should be cleaned every day after use and immediately after spills occur