background image

8

Packing

Clamp Bolt

Support Tube
and Flange
(min. bore
42.0 mm dia.)

Thermal
Insulation

Fixing
Flange

Furnace
Wall

150mm (5.91 in.)
minimum

Min. 42.0 mm dia.

Thermal Insulation

Furnace
Wall

Soft Ceramic Fibre

Support Tube

10

°

 min.

Thermal
Insulation

Support
Tube

Furnace
Wall

Soft
Ceramic
Fibre

…3

INSTALLATION

3.3

Mounting – Figs. 3.7 to 3.9

The probe may be fitted horizontally or, preferably, vertically
for prolonged high temperature operation.

Mount the probe through the furnace wall using the preferred
method shown in Fig. 3.7. Ensure that a sufficiently large hole
is provided through the mounting (min. bore 42mm diameter)
to avoid damage to the probe during insertion or use.

Drill the fixing flange to suit the local fixing requirements before
assembling the probe – see Section 2.3

.

Caution

. To prevent furnace gas from

contaminating the reference air, always ensure that the
clamp ring, fixing flange and sealing ring are fitted to the
probe as described in Section 2.3.

In applications such as high temperature incinerators, fluid
bed boilers, and ore roasters, where the level of water vapour
in the waste gases to be measured can be extremely high, it is
important that the mounting flange of the probe, including any
stand-off which may be used, is thermally insulated to
minimize condensation within the probe – see Fig. 3.7.
Condensation within the probe sheath, particularly in outdoor
installations, can be sufficient to allow water to come into
contact with the hot ceramic (zirconia) tube leading to thermal
shock and failure of the sensor. This problem is most likely to
occur on such installations where the plant is shut-down
regularly over the week-end. Additionally, take care to protect
the head of the probe from the elements in out-door
installations.

Raising the temperature of the probe as rapidly as possible will
assist in reducing the level of condensation. This is not
possible on some processes where the plant temperature is
gradually raised to the normal operating level. The
temperature rise at the head of the probe can be assisted by
having a large clearance hole in the refractory allowing hot
gases to get up to the mounting flange/stand-off – see Fig. 3.8.

The problems described above are more likely to occur on
horizontal installations. Where it is not possible to mount the
probe in the vertical position, install the probe with a slope of at
least 10 to 15

°

 downwards from the head of the probe –

see Fig. 3.9.

Fig. 3.7 Probe Mounting (standard configuration)

Fig. 3.8 Probe Mounting (for improved probe heating)

Fig. 3.9 Probe Mounting (inclined configuration)

Summary of Contents for ZGP2 Series

Page 1: ...AnalyzeIT High Temperature Zirconia Oxygen Probe ZGP2 Series Operating Instructions IM ZGP2_8 ...

Page 2: ...l conditions result in degraded process system performance leading to personal injury or death Therefore comply fully with all Warning and Caution notices Information in this manual is intended only to assist our customers in the efficient operation of our equipment Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without...

Page 3: ...librium shift common to other systems in which gas samples are cooled before measurement and has a fast response to changes in atmosphere enabling rapid corrective action to be taken when necessary The probe comprises a ceramic detector cell housed in a protective sheath A thermocouple is fitted within the probe to enabletheprocesstemperaturetobemonitoredorforautomatic temperature compensation to ...

Page 4: ...h is fitted adjust the diameter of the wadding at the end of the probe to allow a snug fit To reduce its diameter the wadding may be compressed slightly To increase its diameter apply slight end pressure on the wire bonding at both ends of the wadding Mould the wadding by hand to ensure a light push fit into the sheath 2 2 Checking the Code Number Ensure that the correct equipment is being install...

Page 5: ...3 Wadding Filter Washer Vent Holes 1 to 1 5 mm 1 3 2 5 6 7 8 8 3 4 5 8 2 2 PREPARATION Fig 2 1 Assembling the Probe ...

Page 6: ...urement is required i e O2 ppmO2 or O2 partial pressure temperature compensation of the probe output is usually necessary If the process temperature is constant or if variations in operating temperature do not produce unacceptable errors as in some flue gas measurement it is possible to measure O2 concentration without temperature compensation using a system similar to that for oxygen potential Sy...

Page 7: ...uple mV Output from Zirconia Cell Thermocouple Output from Probe Z MT Oxygen Analyzer Alarm Control or Retransmission Outputs Alarm Control or Retransmission Outputs Alarm Control Retransmission or Serial Data Output Power Supply Remote Fuel Selector dual fuel versions only Carbon Monoxide Monitor 3 INSTALLATION Fig 3 2 Oxygen Concentration Measurement using Basic Oxygen Analyzer Fig 3 3 Oxygen Co...

Page 8: ... subsequent removal Dimensions for the probe are shown in Figs 3 4 to 3 6 A clearance of at least 25mm in excess of the overall probe length is necessary for installation or removal procedures Caution The probe MUST NOT be sited a where it is subjected to mechanical or thermal shock b in the presence of aggressive components e g molten slags molten silicates metals and vapours of lead zinc silicon...

Page 9: ...ustomer M16 Cable Entries x2 Reference Air Line Entry Mounting Flange Probe Outer Sheath Test Gas All dimensions in mm inches 3 INSTALLATION 3 2 Siting Figs 3 4 to 3 6 Fig 3 5 Overall Dimensions Twin Gland C95 Head Fig 3 6 Overall Dimensions 1 in NPT Adaptor 69 8 2 74 View A A 15 0 59 Twin Gland Head Shown for Information Only All dimensions in mm inches ...

Page 10: ...e of the probe including any stand off which may be used is thermally insulated to minimize condensation within the probe see Fig 3 7 Condensation within the probe sheath particularly in outdoor installations can be sufficient to allow water to come into contact with the hot ceramic zirconia tube leading to thermal shock and failure of the sensor This problem is most likely to occur on such instal...

Page 11: ...able for use with Pt Pt 13 Rh thermocouples to BS4937 sheathing to BS6746 part no E37 4 3 Electrical Connections Fig 4 2 Make connections as shown in Fig 4 2 and on the label inside the probe head cover The connecting cable must enter the probe head via the bush provided standard head or suitable cable glands twin gland C95 head 4 4 Reference Air Connections Fig 4 2 The reference air connector is ...

Page 12: ...ference Gas P1 Partial Pressure Sample Gas C Cell Constant For oxygen potential measurement the following two expressions relate the oxygen probe output E mV to oxygen potential µ E 10 84µ 40 where µ is in kilocalories or E 2 591µ 40 where µ is in kilojoules 5 2 Range of Operation Fig 5 1 The internal resistance of the cell i e the resistance of the electrolyte between the electrodes decreases app...

Page 13: ...nnections b Remove the screwed plug from the test gas connector see Fig 4 2A c Fit 1 4in i d x 3 8in o d plastic or similar tubing and supply a test gas of known oxygen concentration to the probe at a steady flow rate of 800 to 1000ml min 1 7 to 2 1ft3 hr Allow at least 5 minutes for the system to stabilize before making a measurement d Check that the oxygen concentration measured by the probe sys...

Page 14: ...he limits are given in Table 8 1 for 1 electronic conduction but in practice it may be possible to measure even lower oxygen potential levels without introducing significant errors Electrical Data Cell Constant 0V 2mV Probe accuracy Typically better than 5 of reading System accuracy 2 of test gas reading when calibrated against a certified test gas Cell output impedance Typically 100kΩ 700 C 1292 ...

Page 15: ...umentation pH conductivity and dissolved oxygen transmitters and sensors ammonia nitrate phosphate silica sodium chloride fluoride dissolved oxygen and hydrazine analyzers Zirconia oxygen analyzers katharometers hydrogen purity and purge gas monitors thermal conductivity Customer Support We provide a comprehensive after sales service via a Worldwide Service Organization Contact one of the followin...

Page 16: ... Lockheed Way Carson City NV 89706 USA Tel 1 775 883 4366 Fax 1 775 883 4373 ABB has Sales Customer Support expertise in over 100 countries worldwide www abb com The Company s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice Printed in UK 03 04 ABB 2004 ...

Reviews: