Navigator 500
Sodium analyzer
6 Operation overview – transmitter
32
OI/ASO550–EN Rev. G
Fig. 6.5 shows an overview of
Operator Pages 2 to 5
. Each
Operator Page
displays the process value (PV) and temperature from a
single wet-section. Fixed, color-coded, user-assignable tags (one for each connected wet-section) and color-coded bargraphs aid
identification of each wet-section.
The bargraph indicates the PV (minimum and maximum values on the chart are configurable in the
Sensor Setup
level). If the PV is
above the maximum specified range of the wet-section, the bargraph flashes to indicate the value is out of range.
When multiple wet-sections are connected, and if
Autoscroll
is selected from the
Operator Menu
(see Fig. 6.2, page 30), the display
scrolls through each available
Operator
Page
consecutively.
Fig. 6.5 Operator Pages 2 to 5 – overview
ppb
°C
21.7
21.7
CAL
2012–04–15
10 : 34: 52
4.28
4.28
Sodium
Process value (PV)
Unit of measure
Diagnostic icon /
diagnostic message
Temperature /
temperature unit
Cal shortcut prompt
Time / date
Color-coded wet
section bargraph
wet-section 1 = red
wet-section 2 = green
wet-section 3 = blue
wet-section 4 = violet
Product tag
(user-set)
Operator Menu
icon
Color-coded
wet-section tag
wet-section 1 = red
wet-section 2 = green
wet-section 3 = blue
wet-section 4 = violet
Autoscroll
icon
Parameter ID
LLDO
Sodium
Hydrazine
Tag 1
In Configuration
Navigator