5 Calibration
22
OI/ASO550–EN Rev. G
5.2 Performing a calibration for the first time – single-stream wet-section
To perform a quick two-point calibration from an
Operator
page:
1. Press the
key (below the
Cal
prompt).
The
Calibrate
page is displayed:
2. Press the
key (below the
Select
prompt).
The
Calibrate
/
Sodium
1 (2, 3, 4) page is displayed.
Use the
/
keys to select the wet-section to be
calibrated.
3. Press the
key (below the
Select
prompt on the
Calibrate
/
Sodium
1 [2, 3, 4] page).
The
Sodium 1 (2, 3 or 4) / Start Calibration
page
is displayed:
4. Press the
key (below the
Yes
prompt on the
Start
Calibration
page).
The
Calibration
page is displayed with a bar graph to
indicate calibration progress:
At the end of the calibration, and if the calibration is
successful, the
Active Slope
,
Active Offset, Last Slope
and
Last Offset
values in the
Signals View
are updated – see
Section 6.6, page 33 for details of
View
modes.
5. To exit the
Calibration
page, press the
key (below the
OK
prompt). Calibration continues and calibration
progress can be monitored from the
Calibration View
–
see Fig. 5.1, page 24.
The
Calibrate
/
Sodium
page is displayed.
6. Press the
key (below the
Back
prompt).
The
Calibrate
page is displayed:
7. Press the
key (below the
Exit
prompt) to return to the
Operator
page.
CAL
2012–04–15
10 : 34: 52
°C
21.7
21.7
2.93
Navigator
Tag 1
Na
ppb
Menu
Calibrate
Exit
Select
Sodium 1
Calibrate
Back
Select
Sodium 1
Start Calibration
Exit
Yes
Note.
The
Active Slope
and
Active Offset
values are
not updated if the calibration fails. The
Calibration Log
is updated with the
Active Slope
and
Active Offset
values.