23
Test
Test Mode
Test Flow
Test %
Test Hz
Test mA
Test Vel
Test Alm
Test Txv
Test Quit
C
1 >
2 >
3 >
4 >
5 >
6 >
7 >
8 >
Q
C >
Sub-menus
Remarks
Main Menu
System Test: Exercises all outputs (as programmed) from manually entered flowrate.
If set to '1', transmitter is in 'Test Mode'
Displays current flowrate: if in 'Test Mode', then any value may be
entered manually
Flowrate as a percentage.
Output Frequency.
Output Current.
Flow velocity in sensor (for the flowrate above).
Shows currently active alarms, sequentially
('Clr' indicates no alarms active).
Live flow velocity (uncorrected for sensor calibration).
Enter Q (CR) to return to Main Menu or (CR) to cycle around this menu.
*
#
If the sensor is empty or disconnected, the alarms 'MtSnsr' and Coil will be displayed as appropriate.
The maximum which can be entered must not exceed 21000. The value entered may be displayed with a small error in the
decimal digits e.g. 1.900 may be displayed as 1.899. This is a display characteristic and the value 1.900 will be used by the
MagMaster.
*
END OF PARAMETERS
Calculated from
'Test Flow 2' above
#
…B1.2
Description of Parameters
✶
Information.
'Test
Mode' self-cancels after 30 minutes, if no further entries are made.