To Calibrate the Isotope Values
1.
Connect the WVISS to the inlet of the analyzer.
2.
Use a Water Standard with known isotopic values.
3. If the WVISS is labeled with a Calibration Voltage Sticker (example 2.2V =
10,000ppm), set the manual hand pad to that voltage.
a. If the WVISS is not labeled with a Calibration Voltage, set the hand pad
between 2.2-2.4 Volts and adjust accordingly until the analyzer is reporting
10,000ppm.
4. Allow the analyzer to stabilize at 10,000ppm before calibrating.
5. Click the Setup button on the
User Interface Control Bar. (Figure 91)
Figure 91: User Interface Control Bar
6. Select the Calibration tab at the top of the screen to enter the
Calibration menu.
(Figure 92)
Figure 92: Calibration menu