Figure 86: Gas Manifold Control Screen for the MIU, Enabled
If a valve is set to 0, the entry is ignored. Each defined gas is sampled
sequentially in its respective group (unknown or reference).
4. Populate the reference gas valve sequence:
a. Valve - The current valve being referenced. Corresponds to the port number
on the MIU.
b. Seconds - How long the analyzer should reference the gas (in seconds).
c.
Description- Input a short text description associated with the reference
gas connected to that valve.
5. Click on the Start with reference gas valve sequence check box if you wish to run
your reference gases first.
6. Use the arrow scroll bar to select the number of times to run the unknown gas
sequence for each reference gas sequence. (Figure 86)
7.
Select Save Changes to save your current configuration.
8. To begin sampling, click Close. (Figure 86)