USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
95
10. Each step is displayed in the lower-right panel of the calibration screen as the
analyzer performs the calibration. Figure 90 shows the calibration process as a
flow chart.
Figure 90: Calibration Flow
11. When the
Calibration Complete message is displayed, click the CLOSE button.
12. Enter
TimeChart by selecting the Display button on the User Interface Control Bar
(Figure 88) and verify that the displayed concentration is reporting the expected
value.
The time of latest calibration is stored in
Reference Gas Settings within
the
Calibration menu for future reference.