USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
108
Configuring the WVISS and MIU to run simultaneously
The MIU and WVISS accessories can be configured together to sample multi-channel
measurements along with automatic water vapor isotope references.
Connect the Components
Connect the accessories to the analyzer in this order: Samples
MIU
WVISS
Analyzer
The tubing connecting the components should be as short as possible to
minimize the response time when switching channels.
1.
Connect your samples to the MIU inlet ports. Refer to the
MIU Appendix for setup.
2.
Connect a ¼” tube from the MIU Outlet push-connect port to the WVISS Sample In
push-connect port on the front panel of the WVISS.
3. Connect the WVISS Outlet port to the analyzer Inlet port.
Configure the WVISS and MIU Setup Panels
1.
Setup the MIU screen. Refer to the
MIU Appendix for details.
a. Click on the MIU tab at the top of the
Setup menu selection bar to access
the
MIU Setup Panel.
Figure 108: MIU Setup Panel