Parameterization
D184B140U02 FSM4000
87
Submenu/parameter
Value range
Comment
Submenu
Syst. Adjust
Start
System
Zero
Manual
Autom. Calibration
Check zero, if necessary (when adapting an older
flowmeter sensor). Manual entry, e.g., when replacing the
transmitter.
0.0
%
System
zero
point Numerical
entry
Automatic adjustment: Valve must be closed, fluid in
flowmeter sensor must be at absolute standstill. Press
ENTER to perform the automatic adjustment. The limit for
the zero point is ± 10%. If the value is outside this limit, no
calibration is performed.
0.0
%
FSM4000 01/2010
Display only
Display software version. 01.2010 specifies the date of
the release. C0.11 = Version level, for standard and
HART protocol models.
D699G004U01
C0.11
FSM4000
PA3.0
Display
only
For
PROFIBUS
PA
software.
D699G004U02
C0.13
FSM4000
FF
Display
only
For FOUNDATION Fieldbus communication software.
D699G004U03
C0.14