Parameterization
D184B140U02 FSM4000
63
Submenu/parameter
Value range
Comment
Language
German
English
French
Finnish
Spanish
Italian
Dutch
Danish
Swedish
Turkish
The languages available are displayed in the present
language selection. Menus in the Service Code section
are displayed in English.
English
Language
English
Submenu
Operating mode
Operating
mode
Standard
Fast
Piston Pump
"Standard": Selection used for continuous flow metering.
"Fast": Accelerated processing of flowrate (e.g., short
batch operation).
"Piston Pump": For pulsating flows. In this operating
mode, data logging improves reproducibility for pulsating
flows and during pump operation. The totalizer is
automatically set for the operating mode Difference
Totalizer. Any reverse flow that may occur is
automatically deducted from the totalizer value.
Standard
Flow
indication
Standard,
inverse Standard/inverse, reverse the flow direction indicators in
the display for forward and reverse flow.
Standard
Noise
reduction
Off, 1, 2, 3, 4
The noise reduction is switched on when the measured
value is:
-
very unstable or
-
temporarily or permanently frozen (error E)
Note: Activating noise reduction increases the response
time.
Off
Flow direction
Forward/Reverse
Forward
When the flow indication "Forward/Reverse" is selected,
the flow direction is indicated in the display by "
→
F" for
forward and "
←
R" for reverse. If "Forward" is selected,
the totalizer only totalizes flow in the forward direction. If
the flow is reversed, warning 10 "Reverse Q" is displayed.
Forward/Reverse
Submenu
Unit
Unit
totalizer
ml; l; m
3
; ugl; igl; bbl;
g; kg; t; lb;
Prog Unit
1)
; ft3
2)
Select the engineering units for volume and mass flowrate
totalization. If these units are changed, the transmitter
checks and updates the flow range, pulse factor and
pulse width (0.1 ms ... 2000 ms). The transmitter first
changes the pulse factor and then the pulse width.
1) kgal = Factory setting
2) ft3 = added only for PROFIBUS PA
ml