E N DU R A A Z 20, A Z 2 5, A Z 3 0

  |  P U M P ED R EFER EN C E A I R U N I T  |  I N F0 9/0 2 7- EN R E V. F

Electrical Safety – CEI / IEC 61010-1:2001-2

This equipment complies with the requirements of CEI / IEC 
61010-1:2001-2 ‘Safety Requirements for Electrical Equipment 
for Measurement, Control and Laboratory Use’ and complies 
with US NEC 500, NIST and OSHA. 

If the equipment is used in a manner NOT specified by the 
Company, the protection provided by the equipment may be 
impaired.

 WARNING

•  Isolate electrical supplies to the system and reference air 

pump enclosure before proceeding with  pump 
replacement.

•  Isolate the air supply to the reference air pump (part 

numbers AZ200 770/AZ200 771/AZ200 772/AZ200 773) 
before replacing the pump unit.

•  Observe all relevant health and safety procedures for 

equipment maintenance.

Product recycling and disposal 

(Europe only)

  Electrical equipment marked with this symbol may 

not be disposed of in European public disposal 
systems after 12 August 2005. To conform to 
European local and national regulations (EU Directive 
2002/96/EC), European electrical equipment users 
must now return old or end-of-life equipment to the 
manufacturer for disposal at no charge to the user. 
ABB is committed to ensuring that the risk of any 
environmental damage or pollution caused by any of 
its products is minimized as far as possible.

Information on ROHS Directive 2011/65/EU 

(RoHS II)

  ABB, Industrial Automation, Measurement & Analytics, 

UK, fully supports the objectives of the ROHS II 
directive. All in-scope products placed on the market 
by IAMA UK on and following the 22nd of July 2017 and 
without any specific exemption, will be compliant to 
the ROHS II directive, 2011/65/EU.

Service and repairs

Other than the serviceable items listed in this document, none 
of the pump’s components can be serviced by the user. Only 
personnel from ABB or its approved representative(s) is (are) 
authorized to attempt repairs to the system and only 
components formally approved by the manufacturer should be 
used. Any attempt at repairing the pump in contravention of 
these principles could cause damage to the pump and corporal 
injury to the person carrying out the repair. It renders the 
warranty null and void and could compromise the correct 
working of the pump and the electrical integrity or the CE 
compliance of the pump. 

If you have any problems with installation, starting, or using the 
pump please contact the company that sold it to you. If this is 
not possible, or if the results of this approach are not 
satisfactory, please contact the manufacturer’s Customer 
Service

Potential safety hazards 

The following potential safety hazards are associated with 
operating the system: 

•  Electrical (line voltage)

Summary of Contents for Endura AZ20 series

Page 1: ...AC 50 60 Hz AZ200 770 AZ200 770 1 4 BSP Metric 115 V AC 50 60 Hz AZ200 771 AZ200 771 1 4 NPT Imperial 230 V AC 50 60 Hz AZ200 772 AZ200 772 1 4 NPT Imperial 115 V AC 50 60 Hz AZ200 773 Table 1 ABB pu...

Page 2: ...s useful or important information about the product Product symbols Symbols that may appear on this product are shown below Protective earth ground terminal Functional earth ground terminal This symbo...

Page 3: ...age or pollution caused by any of its products is minimized as far as possible Information on ROHS Directive 2011 65 EU RoHS II ABB Industrial Automation Measurement Analytics UK fully supports the ob...

Page 4: ...re cable rated 5A and 90 C 194 F minimum that conform to either IEC 60227 or IEC 60245 The terminals accept cables from 0 8 to 2 5 mm2 18 to 14 AWG The pump conforms to Installation Category II of IEC...

Page 5: ...4 make mains connections as detailed in steps 1 to 5 The mains power supply must be 115 or 230 V AC 50 60 Hz depending on the pump model refer to Table 1 page 1 1 Prepare the incoming power supply cab...

Page 6: ...ion inlet A and ensure the other end of the pipe is located in a clean dry area 2 Connect the pumped reference air outlet B directly to the reference air inlet on the probe refer to individual probe m...

Page 7: ...eplacing the air filter 1 Remove the pump cover see Accessing the terminal block page 5 Referring to Figure 6 2 Disconnect tube A from air filter B 3 Disconnect tube C from air filter B 4 Release clip...

Page 8: ...ghts in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden withou...

Reviews: