Combustion oxygen monitor
8 Troubleshooting
82
IM/AZ20E–EN Rev. F
8.2.2 Out of Specification Messages
Message range: S068.040 to S048.005
Status
Icon
Diagnostic Message
Displayed
Possible Cause(s)
Corrective Measure(s)
––S068.040––
INTERNAL TEMP.
OUT OF RANGE
Check internal
temperature
Ambient temperature too high /
low or fault in electronics.
Modify ambient temperature.
Replace cartridge.
If problem persists contact local
ABB service.
––S064.032––
CAL. FACTOR
FAILED
Cell is faulty.
Replace cell.
Cell is faulty.
Replace cell.
––S062.033––
CAL. OFFSET
FAILED
Cell is faulty.
Replace cell.
Cell is faulty.
Replace cell.
––S060.028––
CALIBRATION
STABILITY FAILED
Check test gas
connections.
Re-calibrate
Cell input measurement is too
noisy.
Check gas connections.
Repeat calibration.
If problem persists contact local
ABB service.
––S058.019––
MISSED SCHEDULED
AUTOCAL
Check reason for
missed cal.
Re-calibrate
Device not in operating mode
when calibration was scheduled.
Check reason why automatic
calibration was not performed.
Perform calibration.
––S056.016––
MAINS FREQUENCY
ERROR
Check mains
power source.
Power frequency not within 45 Hz
to 65 Hz range.
Check power source.
Table 8.3 Diagnostics – Out of Specification Messages