Endura AZ series integral and remote transmitter
Combustion oxygen monitor
5 Programming
IM/AZ20E–EN Rev. F
33
5.4.1 Easy Setup
The
Easy Setup
menu contains a series of easy setup options
for users with
Advanced
level access.
Standard
and
Read Only
level users cannot access the
Easy
Setup
menu.
Parameter
Comment / [Range]
Default
Language
Selects the language displayed on the transmitter's local
display – see Section 5.4.5, page 48 for language options.
English
Instrument Tag
Used to enter an instrument tag (up to 20 characters)
displayed on the
Operator Page(s)
Characters are selected from an alphanumeric list – see
Section 5.1, page 26 for navigation details.
The tag is not displayed on the
Operator Page(s)
if a
Display Mode
of
3 x 9
is selected – see Section 5.4.5, page
N / A
Date Format
Selects the format of the displayed date.
DD-MM-YYYY
MM-DD-YYYY
YYYY-MM-DD
YYYY-MM-DD
Time & Date
Sets the current local time and date.
N / A
Hr:Min:Sec
Year:Month:Day
Probe Type
Selects the probe type to be used with the transmitter.
AZ20 &
Remote Tx*
AZ20 & Integral Tx
AZ20 & Remote Tx
AZ25 & Remote Tx
AZ30 & Integral Tx
AZ30 & Remote Tx
AZ35 & Remote Tx
AZ10 With Filter
AZ10 No Filter
*For marine application Exhaust Gas Recirculation (EGR) transmitters, the probe type is set to
AZ10 No
Filter
and cannot be changed.
Table 5.6 Easy Setup Menu
1
Menu
Exit
Select
Easy Setup