7 Operator Pages and Menus
OI/APA592–EN Rev. E
63
7.3.3 Device Setup
Used to set the sensor / sensor input, run temperature sensor
recognition and obtain transmitter information.
Note
. Required access level =
Advanced
.
Device Setup Parameters
Parameter
Description
Tx
Type
Sensor Type
Analyzer Type
Sets the analyzer type:
pH
ORP
pION
Ion Conc
All
pH Sensor Type
Selects the type of sensor used for measurement:
Glass
Antimony
Custom
pH
Isopotential Pt.
(Isopotential Point)
Enabled when
pH Sensor Type
=
Custom
.
Adjustable range –20 to 20 pH. Default = 7 pH.
Enabled when
pH Sensor Type
=
Custom
.
Asymmetric Pt
(Asymmetric Potential)
Adjustable range –2000 to 2000 mV. Default = 0 mV
Enabled when
pH Sensor Type
=
Custom
.
Valence
Ion valence selection of either –3, –2, –1, 1, 2, or 3.
The valence determines the millivolt change per decade of concentration.
Ion
Conc
Magnitudes
The number of magnitudes ranging from 1 to 3 that defines the ion
concentration output.
End Mag
The ion concentration state functions by associating an end millivolt value to
an end magnitude value. The end magnitude value can be set to 10, 100 or
1000.
End mV
The end millivolt value associated with the end magnitude value.
Sensor Diag
Turns sensor diagnostics on (Yes) or off (
No
).
All
Reference Imp. Limit
A reference impedance value above this limit triggers a diagnostic error for
high-reference impedance.
Adjustable range 1 to 1000 K
. Default = 50 K
.
Limit enabled when
Sensor Diag
=
Yes
.
3
Menu
Exit
Select
Device Setup