7 Operator Pages and Menus
62
OI/APA592–EN Rev. E
Temperature Calibration
6
At the calibration results screen(s) check
the new values.
Press the
key to exit
the screen:
or
if a calibration failure message is
displayed, reset the buffer values (see
steps 2 to 5).
7
If the
Hold Output
screen was selected at
step 1, an option to release the output is
provided at the
Release Output
screen.
Press the
key to retain the
Hold
Output
state or the
key
release the
Hold Output
state.
Step
Procedure
Screen
1
At the
Temperature Cal
parameter, wait
for the message '
Continue When Stable
'
to be displayed then press the
key to
continue.
2
Enter the new temperature using the
and
keys to select the new value then
press the
key to continue display the
result screen.
Table 7.5 Temperature Calibration
Step
Procedure
Screen
Table 7.4 Auto Buffer Calibration (Continued)