EL6010 Gas Analyzer in Category 2G Operator’s Manual
39
Perform Calibration Reset
When should a
calibration reset be
performed?
Perform a calibration reset when the gas analyzer can no longer be calibrated with
normal means. A possible cause of this is calibration with the wrong test gases.
What does the
calibration reset do?
A calibration reset returns the gas analyzer’s calibration to factory default values.
Additionally, the offset drift and amplification drift are electronically returned to
initial calibration values (see Section “Drift Reading”, page 42).
Step Action
Key Display
Key
1
Enter
Calibrate
Enter
Select the Calibrate
menu.
2
×
Ø
Calibrate component
CO
Enter
Select the sample
component.
3
×
Ø
Calibration reset
CO
Enter
Select Calibration reset.
4
Calibration reset
Start cal. reset!
Enter
Start the calibration
reset.
5
Wait cal. is running
Calibration reset in
progress.
6
Calibration reset
CO
Calibration reset has
ended.
Perform
Calibration Reset
7 Press
the
Back
key to
return to displaying
measured values.