EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Power Supply Connection
Step Action
1
Make sure that the power line voltage is in the permissible range
85 to 250 VAC.
Make sure the power supply leads have an adequately dimensioned
protective device (max. 6 A breaker).
2
Power consumption:
EL6010-CU + EL6010-Uras14
approx. 100 VA
EL6010-CU + EL6010-Magnos106
approx. 90 VA
EL6010-CU + EL6010-Caldos17
approx. 40 VA
3
Install a breaker in the power supply line or a switched receptacle near
the control unit to make sure the control unit can be completely sepa-
rated from the power source. Mark the breaker so that its relationship
to the protected device is clear.
Connecting the
Power Supply to the
Control Unit
4
Connect the power supply leads (conductor cross-section = 1.5 to
2.5 mm
2
) to the terminals N, L and PE (see connection diagram,
Analyzer Unit
Power Supply
The analyzer unit is supplied with 24 VDC
±
5 % from the control unit’s power
supply.