EL6010 Gas Analyzer in Category 2G Operator’s Manual
42/24-30 EN Rev. 1
Limit Parameters
Limit Parameters
If only one sample component is present in the gas analyzer, two limits can be set
for this component.
If there are two sample components in the gas analyzer, one limit can be set per
sample component.
Switching hysteresis can be set for each of the limits.
Limit 1
CO
OFF > 400.0 ppm
Limit of the sample component displayed
Limit 1 CO
OFF
> 400.0 ppm
Activation.
Value range: ON or OFF
Limit 1 CO
OFF
>
400.0 ppm
Direction of effect.
Value range: < or > = Alarm on underflow or
overflow of limit
Limit
Limit 1 CO
OFF >
400.0
ppm
Limit.
Value range: Within the reference measuring range
1)
Step Interval
The limit adjustment step interval is 0.1 % of the span of the reference measuring
range
1)
.
Limit 1 hysteresis
10.0 ppm
Switching Hysteresis
Switching Hysteresis
Limit 1 hysteresis
10.0
ppm
Extent of switching hysteresis.
Value range: Within the reference measuring range
1)
Step Interval
The hysteresis adjustment step interval is 0.01 % of the span of the reference
measuring range
1)
.
1) For an explanation of the term “Reference Measuring Range” see pages