![ABB CoreSense User Manual Download Page 21](http://html.mh-extra.com/html/abb/coresense/coresense_user-manual_125021.webp)
Installing the sensor head
17
5
Once fully tightened, make sure that the sensor is in horizontal position with the sampling access
panel pointing upwards. Rotate the sensor as required to achieve this.
NOTICE
Failure to position the CoreSense horizontally, with the access panel pointing upwards,
will cause the thermal pump to malfunction and may result in erroneous readings.
6
With the 5 mm Allen key, tighten the rotating flange nuts to a torque of 8.1 N-m (6.0 lb-ft) according
to the following tightening pattern.
5
6
3
4
2
7
1
8
Start running in all the bolts or nuts a few threads and then run them hand-tight. Tighten by hand,
then according to the torque specification.
NOTICE
Do not over tighten.
Summary of Contents for CoreSense
Page 1: ... USER GUIDE CoreSense Hydrogen and Moisture Sensor ...
Page 10: ...Page intentionally left blank ...
Page 32: ...Page intentionally left blank ...
Page 43: ... Configuring the sensor head 39 About dialog box Figure 23 About dialog box ...
Page 48: ...Page intentionally left blank ...
Page 58: ...Page intentionally left blank ...
Page 80: ...Page intentionally left blank ...
Page 84: ...Page intentionally left blank ...
Page 106: ...Page intentionally left blank ...
Page 108: ...Page intentionally left blank ...
Page 110: ...Page intentionally left blank ...
Page 114: ...Page intentionally left blank ...
Page 120: ...Page intentionally left blank ...
Page 121: ......