INSTALLATION

2.2.4 System Setup (Fig. 2.6)

Warning. Care is needed when

CalMaster is used with cathodically protected
MagMasters.

Cathodically protected installations, wired as

directed in the MagMaster manual, which
isolate the sensor using earthing flanges, are
electrically earthed and do not pose any safety
issues.

When apparatus, such as CalMaster or the

PC, are connected to systems which 'float' the
sensor at the cathodic potential, this apparatus
will also float at the cathodic potential and
therefore must not be grounded or
accidentally grounded.

With the adaptor PCB installed in the MagMaster
Transmitter as detailed in Section 2.2.2, and the
sensor cable plugged into the adaptor as for
normal operation:

Fig. 2.6 Connection Details for Initialisation

MagMaster

CalMaster

1 2 3 4

Sensor

Sensor
Test Lead Set

ASEA BROWN BOVERI

CalMaster

a) Connect lead 3 of the CalMaster to the

MagMaster 9-pin D-Type socket. The three
small probes should not be connected until
requested to do so by the CalMaster software.
For details of when to connect these probe
leads, follow the instructions on the provided
software – see Section 3.2. For details of
where to connect the probe leads, follow the
'on-line help' under the heading, 'test lead
connections'.

b) Connect the 9-Pin D-Type socket on the

CalMaster to a computer (Serial Comms
Connection).

Note.The use of CalMaster with a line

powered desktop PC, as opposed to a
portable, battery powered PC, is not
recommended, as the resulting ground (eqrth)
loop can corrupt the sensitive CalMaster
measurements.

In these circumstances, an external RS232

isolator between the PC and the CalMaster will
effect a solution

Summary of Contents for CalMaster

Page 1: ...ABB Instrumentation Instruction Manual CalMaster CalMaster ASE A BRO WN BOV ERI ...

Page 2: ...ssibility of an accident occurring when operating in conditions of high pressure and or temperature 5 Chemicals must be stored away from heat protected from temperature extremes and powders kept dry Normal safe handling procedures must be used 6 When disposing of chemicals ensure that no two chemicals are mixed Safety advice concerning the use of the equipment described in this manual or any relev...

Page 3: ...2 Preparing the MagMaster Transmitter for Verification 3 2 2 1 Removing the Covers 3 2 2 2 Fitting the Adaptor Board 4 2 2 3 Wiring the Adaptor Socket 5 2 2 4 System Setup 6 3 OPERATION 7 3 1 Software Installation 7 3 1 1 Windows 3 1 or 3 11 7 3 1 2 Windows 95 7 3 2 Initialising the CalMaster 7 3 3 Battery Charging 7 ...

Page 4: ...nsor wiring when the testing is complete 1 PREPARATION 2 INSTALLATION 1 1 Unpacking Unpack and visually inspect the CalMaster Also packed with the CalMaster are a CalMaster sensor test lead set b CalMaster to MagMaster communication lead set includes three test probes for frequency measurement c Adaptor PCBs incorporating a plug to accept the socket in item d below d Socket which fits into item c ...

Page 5: ...the MagMaster Transmitter for Verification 2 2 1 Removing the Covers 2 INSTALLATION Fig 2 2 Access to fit Adaptor Board 2 2 and Slide Off 1Slide Down 4 Slacken Captive Screws Remove Protection Cover 3 Pull Out Slightly ...

Page 6: ...o the terminals previously occupied by the sensor wiring see Fig 2 4 d Fit and tighten the securing screw to the corner of the adaptor board e Tighten the associated terminal block screws Fig 2 3 Normal Sensor Wiring in the Transmitter Fig 2 4 Fitting the Adaptor Board DS 2 IC 2 CD 1 CD 2 SIG GND DS 1 SIG 1 Grey Coax Pink Inner SIG 2 Screen DS 2 Red CD 1 Yellow CD 2 White Coax Blue Inner SIG 1 Scr...

Page 7: ...the screens of the grey and white coaxial wires 3 Insert the adaptor socket into the plug on the adaptor PCB Warning Testing the MagMaster involves operation with the MagMaster terminal cover removed exposing live terminals Take precautions to avoid electric shock hazard Switch on the power to the MagMaster and check that the system operates correctly If the MagMaster operates correctly either the...

Page 8: ...2 6 Connection Details for Initialisation MagMaster CalMaster 1 2 3 4 Sensor Sensor Test Lead Set ASEA BROWN BOVERI CalMaster a Connect lead 3 of the CalMaster to the MagMaster 9 pin D Type socket The three small probes should not be connected until requestedtodosobytheCalMastersoftware For details of when to connect these probe leads follow the instructions on the provided software see Section 3 ...

Page 9: ...follow the instructions 3 2 Initialising the CalMaster Note It is advisable to avoid the use of radio equipment in the vicinity of a CalMaster MagMaster test setup during testing a FromWindows ProgramManager openthe ABB Applications folder and double click on the CalMaster icon The program automaticallyfindstheMagMasterandopens up to the Main ABB Screen d Select Test Meter and the Verification Inf...

Page 10: ...8 NOTES ...

Page 11: ...ure I to P Converters and Field Indicators Flow Metering Magnetic Flowmeters Electromagnetic insertion type probes and watermeters Turbine Flowmeters Wedge Flow Elements Mass Flow Meters Transmitters sensors controllers and batch display units Level Control Submersible Capacitance Conductivity Pneumatic Instrumentation Transmitters Indicating Controllers Recording Controllers Customer Support ABB ...

Page 12: ...AS Issue 2 ABB Kent Taylor Ltd St Neots Cambs England PE19 3EU Tel 01480 475321 Fax 01480 217948 ABB Kent Taylor SpA 22016 Lenno Como Italy Tel 0344 58111 Fax 0344 56278 ABB Instrumentation Inc PO Box 20550 Rochester New York 14602 0550 USA Tel 716 292 6050 Fax 716 273 6207 ABB Kent Taylor Ltd Analytical Flow Group Stonehouse Glos England GL10 3TA Tel 01453 826661 Fax 01453 826358 ...

Reviews: