5
…2
INSTALLATION
2.2.3
Wiring the Adaptor Socket
(Fig. 2.5)
The wires in the sensor cable must now be
wired to the 8-way adaptor socket which will
then connect to the plug on the adaptor
board.
1
Adjust the wire lengths of the sensor
cable to allow for termination and
subsequent removal of the adaptor
socket.
2
Insert the sensor cable wiring into the
socket as shown in Fig 2.5, and tighten
the plug screws with the screwdriver
provided.
Caution. Ensure that no bare
wires touch each other; in particular,
ensure that there is no connection
between the screens of the grey and
white coaxial wires.
3
Insert the adaptor socket into the plug on
the adaptor PCB.
Warning. Testing the MagMaster
involves operation with the MagMaster
terminal cover removed, exposing live
terminals. Take precautions to avoid
electric shock hazard.
Switch on the power to the MagMaster and
check that the system operates correctly.
If the MagMaster operates correctly, either
the transmitter covers can be replaced for
normal working or the MagMaster can be
tested by the CalMaster.
Fig 2.5 Fitting the Sensor Wiring onto the Adaptor Socket
Blue
CD1
CD2
GROUND
3
1
2
Red
Yellow
Pink
Grey
White
Caution. Remove any
exposed black conductive layer
from the inner insulation of both
coaxial cables
DS 2
DS 1
SIG 1
SIG GND
SIG 2