3
68
(2.68)
42
(1.65)
Fixing Centers
160 (6.3)
69 (2.72)
Fixing Centers
Allowance for
Cable Bends
200 (7.9)
61 (23/8) o.d.
Vertical Post
214
(8.43)
232
(9.13)
Three holes Ø6.3
suitable for M6
fasteners
Dimensions in mm (in)
250
(9.84)
Maximum distance
30 metres
C – Within Environmental Limits
55
C
Max
.
–20
C
Min.
B – Within Temperature Limits
A – Maximum Distance of Instrument to Unit
Locate to suit
IP66
IP65
IP54
Environmental cover
Dissolved Oxygen 9438
IP65
Liquid handling panel
Mark fixing centers
(see Fig. 2.3)
Drill suitable
holes
Fix instrument to wall using
suitable fixings
1
2
3
2.1.2
Dissolved Oxygen Flowcell – Fig 2.7
Allow sufficient clearance (200 mm all around) for easy removal
of the flowcell assembly for maintenance when not installed in
the optional enclosure – see Section 2.3.1 for overall dimensions
of units.
Note.
To eliminate the risk of bubbles accumulating at the
sensor, and hence giving erroneous readings, the flowcell
assembly must be mounted vertically.
2.2
Mounting the Instrument
2.2.1
Wall-mounted Instrument – Figs 2.2 to 2.4
Fig. 2.1 Siting Requirements – Instrument/Sensor
Fig. 2.2 Overall Dimensions
Fig. 2.3 Wall Mounting
2
MECHANICAL INSTALLATION
2.1
Siting Requirements
2.1.1
Instruments – Fig. 2.1
Caution.
• Mount instruments in a location free from excessive
vibration.
• Mount away from harmful vapours and/or dripping fluids.
Information.
It is preferable to mount the transmitter at eye
level thus allowing an unrestricted view of the front panel
displays and controls.