1.3 Specifications
Power Requirements
See Section 1.2 Model Number Breakdown
Power Consumption
Refer to Signal Converter instruction manual
50XM1000N for power input requirements
Flowmeter Characteristics
Meter Size/Flow Capacity
See Table 1-1
Span
Factory set at specified range between extremes
listed in Table 1-1; can be field adjusted.
Rangeability
100:1
Minimum Liquid Conductivity
To 0.05
µ
S/cm (depending on liquid velocity and
viscosity)
System Accuracy:
Applies from 2 to 100% of
CAL FACTOR.
1% of rate
RFI Protection
Class 3-abc-0.5% (30 V/m-20 to 1000 MHz) Per
SAMA Standard PMC 33.1-1978. Standard with
integral Converters.
Environmental Limits
Enclosure Classification with
or without Signal Converter
IEC 529 IP65, NEMA 4X
Accidental Submergence
IE 529 IP67, NEMA 4X,
30 feet H
2
O/48 h (9 m H
2
O/48 h)
10D1477 MAGNETIC FLOWMETER INSTRUCTION MANUAL
1-6